PSW-360W |
Model |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Rated Output Voltage |
V |
30 |
80 |
160 |
250 |
800 |
Rated Output Current |
A |
36 |
13.5 |
7.2 |
4.5 |
1.44 |
Rated Output Power |
W |
360 |
360 |
360 |
360 |
360 |
Power Ratio |
— |
3 |
3 |
3.2 |
3.125 |
3.2 |
Constant Voltage Mode |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Line Regulation (*1) |
mV |
18 |
43 |
83 |
128 |
403 |
Load Regulation (*2) |
mV |
20 |
45 |
85 |
130 |
405 |
Ripple and Noise (*3) |
|
p-p (*4) |
mV |
60 |
60 |
60 |
80 |
150 |
|
r.m.s (*5) |
mV |
7 |
7 |
12 |
15 |
30 |
Temperature coefficient |
ppm/°C |
100ppm/°C of rated output voltage, after a 30 minute warm-up. |
Remote sense compensation voltage |
V/wire |
0.6 |
0.6 |
0.6 |
1 |
1 |
Rise Time (*6) |
|
Rated Load |
ms |
50 |
50 |
100 |
100 |
150 |
|
No Load |
ms |
50 |
50 |
100 |
100 |
150 |
Fall Time (*7) |
|
Rated Load |
ms |
50 |
50 |
100 |
150 |
300 |
|
No Load |
ms |
500 |
500 |
1000 |
1200 |
2000 |
Transient response time (*8) |
ms |
1 |
1 |
2 |
2 |
2 |
Constant Current Mode |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Line regulation (*1) |
mA |
41 |
18.5 |
12.2 |
9.5 |
6.44 |
Load regulation (*9) |
mA |
41 |
18.5 |
12.2 |
9.5 |
6.44 |
Ripple and noise |
|
r.m.s (*5 ) |
mA |
72 |
27 |
15 |
10 |
5 |
Temperature coefficient |
ppm/°C |
200ppm/°C of rated output current, after a 30 minute warm-up. |
Protection Function |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Over voltage protection (OVP) |
|
Setting range |
V |
3-33 |
8-88 |
16-176 |
20-275 |
20-880 |
|
Setting accuracy |
|
± (2% of rated output voltage) |
Over current protection (OCP) |
|
Setting range |
A |
3.6-39.6 |
1.35-14.85 |
0.72-7.92 |
0.45-4.95 |
0.144-1.584 |
|
Setting accuracy |
|
± (2% of rated output current) |
Over temperature protection (OTP) |
|
Operation |
|
Turn the output off. |
Low AC input protection (AC-FAIL) |
|
Operation |
|
Turn the output off. |
Power limit (POWER LIMIT) |
|
Operation |
|
Over power limit. |
|
Value (fixed) |
|
Approx. 105% of rated output power |
Front Panel |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Display, 4 digits |
|
Voltage accuracy 0.1% + |
mV |
20 |
20 |
100 |
200 |
400 |
|
Current accuracy 0.1% + |
mA |
40 |
20 |
5 |
5 |
2 |
Programming and Measurement (Interface) |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
Voltage programming accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current programming accuracy 0.1% + |
mA |
30 |
10 |
5 |
5 |
2 |
Voltage programming resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current programming resolution |
mA |
1 |
1 |
1 |
1 |
1 |
Voltage measurement accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current measurement accuracy 0.1% + |
mA |
30 |
10 |
5 |
5 |
2 |
Voltage measurement resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current measurement resolution |
mA |
1 |
1 |
1 |
1 |
1 |
PSW-720W |
Model |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Rated Output Voltage |
V |
30 |
80 |
160 |
250 |
800 |
Rated Output Current |
A |
72 |
27 |
14.4 |
9 |
2.88 |
Rated Output Power |
W |
720 |
720 |
720 |
720 |
720 |
Power Ratio |
— |
3 |
3 |
3.2 |
3.125 |
3.2 |
Constant Voltage Mode |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Line Regulation (*1) |
mV |
18 |
43 |
83 |
128 |
403 |
Load Regulation (*2) |
mV |
20 |
45 |
85 |
130 |
405 |
Ripple and Noise (*3) |
|
p-p (*4) |
mV |
80 |
80 |
80 |
100 |
200 |
|
r.m.s (*5) |
mV |
11 |
11 |
15 |
15 |
30 |
Temperature coefficient |
ppm/°C |
100ppm/°C of rated output voltage, after a 30 minute warm-up. |
Remote sense compensation voltage |
V/wire |
0.6 |
0.6 |
0.6 |
1 |
1 |
Rise Time (*6) |
|
Rated Load |
ms |
50 |
50 |
100 |
100 |
150 |
|
No Load |
ms |
50 |
50 |
100 |
100 |
150 |
Fall Time (*7) |
|
Rated Load |
ms |
50 |
50 |
100 |
150 |
300 |
|
No Load |
ms |
500 |
500 |
1000 |
1200 |
2000 |
Transient response time (*8) |
ms |
1 |
1 |
2 |
2 |
2 |
Constant Current Mode |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Line regulation (*1) |
mA |
77 |
32 |
19.4 |
14 |
7.88 |
Load regulation (*9) |
mA |
77 |
32 |
19.4 |
14 |
7.88 |
Ripple and noise |
|
r.m.s (*5 ) |
mA |
144 |
54 |
30 |
20 |
10 |
Temperature coefficient |
ppm/°C |
200ppm/°C of rated output current, after a 30 minute warm -up. |
Protection Function |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Over voltage protection (OVP) |
|
Setting range |
V |
3-33 |
8-88 |
16-176 |
20-275 |
20-880 |
|
Setting accuracy |
|
± (2% of rated output voltage) |
Over current protection (OCP) |
|
Setting range |
A |
5-79.2 |
2.7-29.7 |
1.44-15.84 |
0.9-9.9 |
0.288-3.168 |
|
Setting accuracy |
|
± (2% of rated output current) |
Over temperature protection (OTP) |
|
Operation |
|
Turn the output off. |
Low AC input protection (AC-FAIL) |
|
Operation |
|
Turn the output off. |
Power limit (POWER LIMIT) |
|
Operation |
|
Over power limit. |
|
Value (fixed) |
|
Approx. 105% of rated output power |
Front Panel |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Display, 4 digits |
|
Voltage accuracy 0.1% + |
mV |
20 |
20 |
100 |
200 |
400 |
|
Current accuracy 0.1% + |
mA |
70 |
40 |
30 |
10 |
4 |
Programming and Measurement (Interface) |
PSW |
30-72 |
80-27 |
160-14.4 |
250-9 |
800-2.88 |
Voltage programming accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current programming accuracy 0.1% + |
mA |
60 |
30 |
15 |
10 |
4 |
Voltage programming resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current programming resolution |
mA |
2 |
2 |
2 |
1 |
1 |
Voltage measurement accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current measurement accuracy 0.1% + |
mA |
60 |
30 |
15 |
10 |
4 |
Voltage measurement resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current measurement resolution |
mA |
2 |
2 |
2 |
1 |
1 |
PSW-1080W |
Model |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Rated Output Voltage |
V |
30 |
80 |
160 |
250 |
800 |
Rated Output Current |
A |
108 |
40.5 |
21.6 |
13.5 |
4.32 |
Rated Output Power |
W |
1080 |
1080 |
1080 |
1080 |
1080 |
Power Ratio |
— |
3 |
3 |
3.2 |
3.125 |
3.2 |
Constant Voltage Mode |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Line Regulation (*1) |
mV |
18 |
43 |
83 |
128 |
403 |
Load Regulation (*2) |
mV |
20 |
45 |
85 |
130 |
405 |
Ripple and Noise (*3) |
|
p-p (*4) |
mV |
100 |
100 |
100 |
120 |
200 |
|
r.m.s (*5) |
mV |
14 |
14 |
20 |
15 |
30 |
Temperature coefficient |
ppm/°C |
100ppm/°C of ra ted output voltage, after a 30 minute warm-up. |
Remote sense compensation voltage |
V/wire |
0.6 |
0.6 |
0.6 |
1 |
1 |
Rise Time (*6) |
|
Rated Load |
ms |
50 |
50 |
100 |
100 |
150 |
|
No Load |
ms |
50 |
50 |
100 |
100 |
150 |
Fall Time (*7) |
|
Rated Load |
ms |
50 |
50 |
100 |
150 |
300 |
|
No Load |
ms |
500 |
500 |
1000 |
1200 |
2000 |
Transient response time (*8) |
ms |
1 |
1 |
2 |
2 |
2 |
Constant Current Mode |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Line regulation (*1) |
mA |
113 |
45.5 |
26.6 |
18.5 |
9.32 |
Load regulation (*9) |
mA |
113 |
45.5 |
26.6 |
18.5 |
9.32 |
Ripple and noise |
|
r.m.s (*5 ) |
mA |
216 |
81 |
45 |
30 |
15 |
Temperature coefficient |
ppm/°C |
200ppm/°C of rated output current, after a 30 minute warm-up . |
Protection Function |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Over voltage protection (OVP) |
|
Setting range |
V |
3-33 |
8-88 |
16-176 |
20-275 |
20-880 |
|
Setting accuracy |
|
± (2% of rated output voltage) |
Over current protection (OCP) |
|
Setting range |
A |
5-118.8 |
4.05-44.55 |
2.16-23.76 |
1.35-14.85 |
0.432-4.752 |
|
Setting accuracy |
|
± (2% of rated output current) |
Over temperature protection (OTP) |
|
Operation |
|
Turn the output off. |
Low AC input protection (AC-FAIL) |
|
Operation |
|
Turn the output off. |
Power limit (POWER LIMIT) |
|
Operation |
|
Over power limit. |
|
Value (fixed) |
|
Approx. 105% of rated output power |
Front Panel |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Display, 4 digits |
|
Voltage accuracy 0.1% + |
mV |
20 |
20 |
100 |
200 |
400 |
|
Current accuracy 0.1% + |
mA |
100 |
50 |
30 |
20 |
6 |
Programming and Measurement (Interface) |
PSW |
30-108 |
80-40.5 |
160-21.6 |
250-13.5 |
800-4.32 |
Voltage programming accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current programming accuracy 0.1% + |
mA |
100 |
40 |
20 |
15 |
6 |
Voltage programming resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current programming resolution |
mA |
3 |
3 |
3 |
1 |
1 |
Voltage measurement accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
Current measurement accuracy 0.1% + |
mA |
100 |
40 |
20 |
15 |
6 |
Voltage measurement resolution |
mV |
1 |
2 |
3 |
5 |
14 |
Current measurement resolution |
mA |
3 |
3 |
3 |
1 |
1 |
Common Specification |
Input Characteristics |
PSW |
30V |
80V |
160V |
250V |
800V |
Nominal input rating |
|
100Vac to 240Vac, 50Hz to 60Hz, single phase |
Input voltage range |
|
85Vac ~ 265Vac |
Input voltage range |
|
47Hz ~ 63Hz |
Maximum input current |
|
100Vac |
A |
360W: 5A, 720W: 10A, 1080W: 15A |
|
200Vac |
A |
360W: 2.5A, 720W: 5A, 1080W: 7.5A |
Inrush current |
A |
< 25A for 360W, < 50A for 720W, < 75A for 1080W |
Maximum input power |
VA |
360W: 500VA, 720W: 1000VA, 1080W: 1500VA |
Power factor |
|
100Vac |
|
0.99 |
|
200Vac |
|
0.97 |
Efficiency |
|
100Vac |
% |
77 |
78 |
79 |
79 |
80 |
|
200Vac |
% |
79 |
80 |
81 |
81 |
82 |
Hold-up time |
|
20ms or greater |
Analog Programming and Monitoring |
PSW |
30V |
80V |
160V |
250V |
800V |
External voltage control output voltage |
|
Accuracy and linearity: ±0.5% of rated output voltage. |
External voltage control output current |
|
Accuracy and linearity: ±1% of rated output current. |
External resistor control output voltage |
|
Accuracy and linearity: ±1.5% of rated output voltage. |
External resistor control output current |
|
Accuracy and linearity: ±1.5% of rated output current. |
Output voltage monitor |
|
Accuracy |
% |
±1 |
±1 |
±1 |
±2 |
±2 |
Output current monitor |
|
Accuracy |
% |
±1 |
±1 |
±1 |
±2 |
±2 |
Shutdown control |
|
Turns the output off with a LOW (0V to 0.5V) or short-circuit. |
Output on/off control |
|
Possible logic selections:
Turn the output on using a LOW (0V to 0.5V) or short-circuit, turn the output off using a HIGH (4.5V to 5V) or open-circuit.
Turn the output on using a HIGH (4.5V to 5V) or open-circuit, turn the output off using a LOW (0V to 0.5V) or short-circuit. |
CV/CC/ALM/PWR ON/OUT ON indicator |
|
Photocoupler open collector output; Maximum voltage 30V, maximum sink current 8mA. |
Series and Parallel Capability |
PSW |
30V |
80V |
160V |
250V |
800V |
Parallel number |
Units |
3 |
3 |
3 |
3 |
3 |
Series Number |
Units |
2 |
2 |
2 |
None |
None |
Front Panel |
Indications |
|
GREEN LED’s:
CV, CC, VSR, ISR, DLY, RMT, 20, 40, 60, 80, 100, %W, W, V, A
RED LED’s: ALM |
Buttons |
|
Function, OVP/OCP, Set, Test, Lock/Local, PWR DSPL, Output |
Knobs |
|
Voltage, Current |
USB port |
|
Type A USB connector |
Interface Capabilities |
USB |
|
TypeA: Host, TypeB: Slave, Speed: 1.1/2.0, USB Class: CDC(Communications Device Class) |
LAN |
|
MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask |
GPIB |
|
Optional: GUG-001 (GPIB to USB Adapter) |
Environmental Conditions |
Operating temperature |
|
0˚C to 50˚C |
Storage temperature |
|
-25˚C to 70˚C |
Operating humidity |
|
20% to 85% RH; No condensation |
Storage humidity |
|
90% RH or less; No condensation |
Altitude |
|
Maximum 2000m |
General Specifications |
Weight (main unit only) |
kg |
Approx. 3kg for 360W,
Approx. 5.3kg for 720W,
Approx. 7.5kg for 1080W |
Dimensions (WxHxD) |
mm3 |
360W: 71×124×350mm,
720W: 142×124×350mm,
1080W: 214×124×350mm |
Cooling |
|
Forced air cooling by internal fan. |
EMC |
|
Complies with the European EMC directive 2004/108 /EC for Class A test and measurement products. |
Safety |
|
Complies with the European Low Voltage Directive 2006 /95/EC and carries the CE-marking. |
Withstand voltage |
|
Between input and chassis: No abnormalities at 1500 Vac for 1 minute. |
Between input and output: No abnormalities at 3000 Vac for 1 minute. |
Between output and chassis:
No abnormalities at 500 Vdc for 1 minute for 30V, 80V, 160V models. |
No abnormalities at 1500 Vdc for 1 minute for 250V, 800V models. |
Insulation resistance |
|
Between input and chassis: 500 Vdc, 100MΩ or more |
Between input and output: 500 Vdc, 100MΩ or more |
Between output and chassis:
500 Vdc, 100MΩ or more for 30V, 80V, 160V and 250V models. |
1000Vdc, 100MΩ or more for 800V models. |
Notes
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
*2: From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10% to 90% of rated output voltage, with rated resistive load.
*7: From 90% to 10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output for a load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input voltage. |